Back to Tools

Carryover Study Calculator

Free carryover (q%) calculator for laboratory analysers, using the high-low sequence method. Enter your results and download a printable report.

Carry Over Performance Verification

Calculate Carry Over (q%) for laboratory instruments

Test Entry Form

q is calculated from H2, L1 and L3. H1 and L2 are recorded as part of the a1, a2, b1, b2, b3 sequence.

Sequence Analysis Chart

Add a test to view the chart

Results Summary

Sr. No.Test NameEquipmentDateH1H2L1L2L3q (%)InterpretationDelete
No test data available. Add a row using the form above.

Interpretation & Reference

Carry over effect is calculated using the equation:

Carry Over = (L1 - L3) / (H2 - L3) × 100

"q" may also be called the 'carry-over ratio', and is sometimes expressed as a percentage value. It implies that the influence of carry-over effects on the analytical results is inversely proportional to the concentration difference of the component to be detected by the analytical procedure in the two specimens.

Carry-over within a particular procedure is of interest. It is dependent on the method, including all reagents applied. Therefore, carry-over effects, rather than carry-over itself, need to be determined. For this purpose, a sequence of specimens, usually with high (a) and low (b) concentrations of the analyte to be measured, are processed — for example a1, a2, b1, b2, b3.

Reference: R. Haeckel, Recommendations for definition and determination of carry over effects, Journal of Clinical Chemistry Vol. 10 No. 04 (Oct–Dec 1988, pp. 181–183)

Carry Over formula: q = (L1 - L3) / (H2 - L3) x 100. Acceptable if |q| < 1.0%

About the carryover study

Carryover happens when part of a high-concentration sample is carried into the next sample, falsely raising a low result. Laboratories check for it when they validate a new analyser or after a change to the sampling system, and it is one of the validation studies assessors look for under ISO 15189:2022.

The study runs a high sample and a low sample in a fixed high-low sequence, repeated several times. The tool compares the low results that follow a high sample with the low results that follow another low sample, and expresses the difference as the carryover percentage (q%).

Enter your lab and equipment details and the measured values, and the tool calculates q% and builds a PDF report. Results that cannot be calculated from the data entered are shown as a dash, not as zero. Your entries are stored in your own browser only.

How to use this tool

  1. Enter the organisation name and the equipment you are testing, so they appear on the report.
  2. Add a test and give it a name. Run a high-concentration sample twice (H1, H2) and then a low-concentration sample three times (L1, L2, L3), in that order, without anything in between.
  3. Type the five results into the sequence for that test. The tool calculates the carryover percentage (q%) by the Haeckel method and plots the run.
  4. Add the other tests on the same analyser and repeat, so one study covers every channel you need.
  5. Download the combined PDF report, sign it and file it with your validation records.

Frequently asked questions

Why does the sequence run the high sample before the low ones?

Carryover moves from a high result into the next sample. The first low sample after the highs (L1) is the one most likely to be affected, while the later lows (L2, L3) show the true low value. Comparing them reveals how much of the high sample was carried over.

Which samples should I use?

Pick a high sample near the top of the measuring range and a low sample near the bottom, ideally patient samples or materials with a similar matrix. The bigger the gap between them, the easier it is to detect carryover.

When should a carryover study be repeated?

When a new analyser is installed, after changes or repairs to the sampling probe or wash system, and whenever unexpected high results start appearing after very high samples. Your laboratory's own procedures may set other triggers.

What if carryover is too high?

Check the probe, the wash solution and the analyser's wash settings, then repeat the study. Until it passes, many laboratories rerun any low sample that follows a very high one.

Which analysers need a carryover study?

Any analyser that samples one specimen after another through the same probe or tubing can carry material over, so the study applies to chemistry, immunoassay and haematology analysers alike. It matters most for tests where a very high result can follow a very low one, as with some hormones and tumour markers.

Is my data sent anywhere?

No. The calculations run in your browser, and your entries are saved only in that browser.

Read the full guide: Carryover study in a medical laboratory: procedure, formula and example. More free calculators are on the Tools page. Preparing your laboratory for accreditation? See our NABL consultancy.

Contact on Whatsapp