Back to Tools

Linearity Study

Master sheet configuration

No parameters yet

Click "Add Parameter" to begin

About the linearity study

A linearity study confirms that a measurement procedure gives results proportional to the true concentration across its reportable range. Laboratories run it when they validate a new method or analyser, and when they need to verify the analytical measurement range that the manufacturer claims.

The usual design prepares a series of dilutions that span the range, measures each one, and compares the measured values with the expected values. This tool takes the expected and measured results, fits the regression line, and reports the equation, the coefficient of determination (R²) and the percentage recovery at each level.

Choose the high sample carefully. It should sit near the top of the range the manufacturer claims, so the dilution series covers as much of that range as possible, and it should behave like a patient sample. Mix every dilution well and run them in one session, so drift between runs is not mistaken for non-linearity.

The report is a PDF you can file with your validation records for ISO 15189:2022 and NABL assessments. Values that cannot be calculated from the data entered are shown as a dash, not as zero. Your entries are stored in your own browser only.

How to use this tool

  1. Enter a parameter name, such as glucose or creatinine, and the highest concentration in your series. The tool works out the expected value at each dilution from that highest value.
  2. Prepare the dilution series from your high sample (1:1, 1:2, 1:4, 1:8, 1:16 and 1:32), run each one on the analyser, and type the observed result next to its dilution.
  3. Read the slope, intercept and R² for the regression line, the recovery at each dilution and the chart of observed against expected values. Each recovery is marked so you can see at a glance which levels need a second look.
  4. Add the next parameter and repeat. Every parameter keeps its own data, so one session can cover a whole analyser panel.
  5. Fill in the lab name, equipment name, serial number and study date, then download the PDF report for one parameter or for all of them, ready to sign and file.

Frequently asked questions

What is the difference between linearity and the analytical measurement range?

Linearity is the property being tested: results rise in proportion to concentration. The analytical measurement range (AMR) is the span of concentrations over which that holds without diluting the sample. A linearity study is how a laboratory verifies the AMR the manufacturer claims, before it reports patient results across that range.

Why does the tool use a two-fold dilution series?

Halving the concentration at each step covers a wide range with few samples and keeps the arithmetic simple: each expected value is the highest value divided by the dilution factor. Start from a sample near the top of the claimed range so the series reaches down towards its low end.

Is a high R² enough to prove linearity?

No. R² shows how closely the points follow a straight line, but a method can have a high R² and still recover too little or too much at the top or bottom of the range. That is why the tool reports the recovery at every dilution alongside the regression, and the report shows both.

What should I do if one dilution fails?

Check the preparation first: pipetting, mixing and the diluent used are the usual causes. Repeat that level before drawing conclusions. If the end of the range still fails, the laboratory may need to report a narrower measuring range or dilute samples that fall above it, as its procedure defines.

What does the recovery at each dilution tell me?

Recovery is the observed value as a percentage of the expected value. A recovery close to 100% means the result at that dilution is where it should be. Recoveries that fall steadily at the top or bottom of the series point to the edge of the range where the method stops being linear.

Should I run each dilution more than once?

The tool takes one observed value per dilution. Many laboratories run each dilution in duplicate or triplicate and enter the mean, which smooths out random error so it is not mistaken for non-linearity. Follow what your own procedure specifies.

Is my data sent anywhere?

No. The calculations run in your browser, and your entries are saved only in that browser so you can come back to them. Clearing the browser's site data removes them.

Read the full guide: What is a linearity study in a medical laboratory?. More free calculators are on the Tools page. Preparing your laboratory for accreditation? See our NABL consultancy.